2.1 Power supply connection
Terminal AC Power supply DC Power supply
Earth Earth
N Neutral 0 V (-)
L Phase 24 V (+)
5
It is important to connect the mains earth to the instruments with AC power supply due to
the presence of a mains filter inside that requires this connection.
2.2 Analog output connection
Terminal
+ mA (positive).
- mA (negative).
The analog output can be either active (that means the receptor must be passive) or
passive (that means the receptor must supply the power for the current loop). It is
recommended to use a receptor with an input resistance of less than 700 to guarantee
correct operation.