Technical Data
• Mounting: Vertical.
• Connection: DN-40 PN-16 DIN 2501 Flange.
R 1½” BSP thead.
(Others available on request).
• Maximum length: AISI-316 ... 6000 mm.
PVC, PTFE ... 2500 mm.
PVC, PTFE ... 6000 mm
(internal AISI-316).
• Precision: ± 2 mm.
• Hysteresis: ± 4 mm.
• Working pressure:
PN-16 AISI-316.
PVC, PTFE with internal AISI-316 tube.
PN-10 PVC, PTFE.
• Liquid temperature:
-20ºC...+150ºC en AISI-316, PTFE.
0ºC...+50ºC PVC.
• Ambient temperature:
-20ºC...+60ºC AISI-316, PTFE.
0ºC...+50ºC PVC.
• Liquid density: From 0,45 to 3 kg/l.
• Liquid viscosity: Maximum 1500 cSt.
• Contacts: Reed 1 A 220 V.
(maximum 9 with minimum separation
of 20 mm)
Conforms with Directive 73 / 23 / CEE
2
No. Piece Materials
LC-.../INOX LC-.../PVC LC-.../PTFE
1 Enclosure
DIN 43650/
Aluminium
DIN 43650/
Aluminium
DIN 43650/
Aluminium
2 Connection AISI-316 PVC PTFE
3 Rod AISI-316 PVC PTFE
4 Float AISI-316 PVC PTFE
5 Magnet Alnico Alnico Alnico
6 Reed switch - - -
7 Float stop AISI-316 PVC PTFE
CONSTRUCTION
OPERATION
Inside the guide tube, reed switches are mounted at the
alarm/control points..
As the float passes the position where a reed switch is
located, the magnetic field from the float actives the switch.
The maximum number of alarm or control points is 9.
Commenti su questo manuale