60 M-1
Variable area
flowmeter
Series
Instructions Manual
C-MI-60M1 Rev.: 0 English Version
Technical Data
Accuracy: ± 2,5% of full scale value. Based on
VDE/VDI 3513
Scales: In l/h, l/min, m³/h, kg/h, %, mm., etc
Scale length: 100 mm ± 5
Mounting: Vertical (rising flow direction)
Total length: 220 mm ± 2
Pipe fittings: 1/4” BSP or NPT female thread
Working pressure: 10 ... 15 bar
Fluid temperature: -10 ... +100ºC
The glass can support a thermal shock of 150 ºC if there is
no internal pressure.
The temperature difference between the interior and exterior
of the glass tube must not exceed 80 ºC
Conforms with the Pressure Equipment Directive 97/23/EC.
This equipment is considered as being a pressure
accessory and NOT a safety accessory as defined in the
97/23/EC directive, Article 1, paragraph 2.1.3.
The following instruction manuals are attached:
60-AMD Limit Switch Instructions Manual
60-AMO Limit Switch Instructions Manual
60-AMH Limit Switch Instructions Manual
!
instrumentation
for fluids
Vedere la pagina 0
1 2 3 4 5 6 7 8

Sommario

Pagina 1 - Instructions Manual

60 M-1 Variable area flowmeter Series Instructions Manual C-MI-60M1 Rev.: 0 English Version Technical Data • Accuracy: ± 2,5% of full scale valu

Pagina 3 - Working Principle

Working Principle The flowmeter consists of a float inside a conical tube. The rising flow pushes the float to an equilibrium point. The area obtained

Pagina 4 - Linea de lectura

RECEPTION The flowmeter is supplied ready for installation and service. The blocking elements that hold the float for transport should be removed bef

Pagina 5

RECEPTION The flowmeter is supplied ready for installation and service. The blocking elements that hold the float for transport should be removed bef

Pagina 7

WARRANTY Tecfluid S.A. GUARANTEES ALL ITS PRODUCTS FOR A PERIOD OF 24 MONTHS, after consignment, against all defects in materials and workmanship. Thi